
DVM1090 GB 3
4. Specifications
Max. accuracy is achieved during a one-year period after calibration. Ideal circumstances require a temperature of 18
to 28°C (64 to 82°F) and a max. relative humidity of 75%.
4.1 General Specifications
Measuring Method Dual-slope integration A/D converter
Sampling Time ±0.4sec.
Display LCD 25mm
Max. Display 1999 points (3 ½ digits)
Polarity Indication “-“ indicates negative polarity
Overrange Indication “1”
Displayed Measuring Unit Unit of electrical capacity
Battery-Low Indication “
”-symbol is displayed
Maximum Altitude 7000ft (2000m)
Max. voltage between terminals and earth 1000V DC or AC
Fuse Protection F 200mA/250V (quick-acting, max. resistance 2
Ω
), F 10A/250V (quick-
acting)
Power Supply 9V-battery (e.g. 6F22)
Operating Temperature 0°C to +40°C (+32 to +104°F)
Storage Temperature -10°C to +50°C (+10 to +122°F)
Dimensions 185 x 84 x 38mm
Weight ±300g (battery included)
4.2 Electrical Specifications
4.2.1. DC Voltage
Range Resolution Accuracy
200mV 0.1mV ±0.5% of rdg + 1 digit
2V 1mV ±0.5% of rdg + 3 digits
20V 10mV ±0.5% of rdg + 3 digits
200V 100mV ±0.5% of rdg + 3 digits
1000V 1V ±0.8% of rdg + 3 digits
Input Impedance : 10M
Ω
Overload Protection : 250V DC or rms AC for the 200mV-range, 1000V DC or 700V rms AC for the ranges 2V to
1000V
Max. Input Voltage : 1000V DC
NOTE
Make sure the test leads are actually touching the circuit when performing measurements in the lower voltage
ranges. The reading will fluctuate if the leads don’t touch the circuit.
Commentaires sur ces manuels