Velleman DVM2000 Manuel d'utilisateur Page 14

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 80
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 13
DVM2000 VELLEMAN
14
DC Current
Range Resolution Accuracy Voltage Drop
500µA 0.01µA ± (0.15% + 15)
102µV/µA
5000µA 0.1µA
± (0.15% + 10)
50mA 1µA
1.5mA/mA
500mA 10µA
5A 0.1mA
± (0.5% + 10) 30mV/A
10A 1mA
Above accuracies guaranteed within the full range.
Resistance
Range Resolution Accuracy
500 0.01 ± (0.1% + 10)
5k 0.1
± (0.1% + 5)
50k 1
500k 10
5M 100 ± (0.1% + 10)
50M 1k ± (0.5% + 10)
Above accuracies guaranteed within the full range.
Capacitance
Range Resolution Accuracy
50nF 0.01nF
± (1% + 5)
500nF 0.1nF
5µF 1nF
50µF 10nF
500µF 0.1µF
± (2% + 5)
5000µF 1µF
Above accuracies for film capacitors or better guaranteed within the full range.
Vue de la page 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 79 80

Commentaires sur ces manuels

Pas de commentaire